A Technique to Minimize Impurity Signal from Blank Rhenium Filaments for Highly Accurate TIMS Measurements of Uranium in Ultra-Trace Levels

Mass Spectrometry Letters / Mass Spectrometry Letters, (P)2233-4203; (E)2093-8950
2010, v.1 no.1, pp.17-20
https://doi.org/10.5478/MSL.2010.1.1.017
Jong Ho Park (Korea Atomic Energy Research Institute)
Inhee Choi (Korea Atomic Energy Research Institute)
Kyuseok Song (Korea Atomic Energy Research Institute)
  • 다운로드 수
  • 조회수

Abstract

As background significantly affects measurement accuracy and a detection limit in determination of the trace amountsof uranium, it is necessary to minimize the impurities in the filaments used for thermal ionization mass spectrometry (TIMS). Wehave varied the degassing condition such as the heating currents and duration times to reduce the backgrounds from the filamentsprepared with zone-refined rhenium tape. The most efficient degassing condition of the heating current and the duration time wasdetermined as 3.5 A and 60 min, respectively. The TIMS measurement combined with the isotope dilution mass spectrometry(IDMS) technique showed that the uranium backgrounds were determined to be in a few fg level from blank rhenium filaments. The background minimized filaments were utilized to measure the uranium isotope ratios of a U030 (NIST) standard sample. The excellent agreement of the measurement with the certified isotope ratios showed that the degassing procedure optimized inthis study efficiently reduced the impurity signals of uranium from blank rhenium filaments to a negligible level.

keywords
Uranium, Background, TIMS, Isotopic Analysis, Safeguard, Isotope Ratio


참고문헌

1

Donohue, D. L. (1998). . J. Alloy Compd, 271-273, 11-.

2

Cohen, A. S. (1992). . Int. J. Mass Spectrom. Ion Process, 116, 71-.

3

Heumann, K. G. (1995). . Analyst, 120, 1291-.

4

Rubin, K. H. (2001). . Chemical Geology, 175, 723-.

5

Richter, S. (2003). . Int. J. Mass Spectrom, 229, 181-.

6

Lee, C. G. (2006). . Jpn. J. Appl. Phys, 45, 294-.

7

Suzuki, D. (2010). . Int. J. Mass Spectrom, 294, 23-.

8

Jeon, Y. S. (2008). . Analytical Science and Technology, 21, 487-.

9

Wolff, J. C. (1996). . Int. J. Mass Spectrom. Ion Process, 156, 67-.

10

Heumann, K. G. (1982). . Int. J. Mass Spectrom. Ion Phys, 45, 87-.

투고일Submission Date
2010-11-01
수정일Revised Date
2010-11-25
게재확정일Accepted Date
2010-11-25